Retention time alignment across batches

Skill HolobiomicsLab/asb-skill-collections/packs/metabolomics/gc-ms/skills/retention-time-alignment-across-batches

Curated, evidence-grounded skill and software-tool collections for scientific AI agents, generated by the AgenticScienceBuilder

Install
npx -y skills add HolobiomicsLab/asb-skill-collections --skill retention-time-alignment-across-batches

Assembled from the repository path, not quoted from the project. Check it against their README if it does not work.

One thing to look at

  • 14 stars14 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.

What its author says it does

Copied from the file, not written here

Use when you have extracted peaks from multiple LC/HRMS batches (n > 500 samples across different analytical runs or days) and observe systematic retention time drift or offset between batches, preventing reliable cross-batch peak matching on m/z and RT alone.

The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.

SKILL.md

8.4 KB, as published. Nobody here has run it

retention-time-alignment-across-batches

Summary

Harmonize retention time values across multiple LC/HRMS batches to align peaks with equivalent m/z and chromatographic profiles, enabling consistent peak-to-compound mapping in population-scale untargeted metabolomics studies. This is essential for multi-batch peak alignment and accurate metabolite quantification across large cohorts.

When to use

Apply this skill when you have extracted peaks from multiple LC/HRMS batches (n > 500 samples across different analytical runs or days) and observe systematic retention time drift or offset between batches, preventing reliable cross-batch peak matching on m/z and RT alone. The skill is triggered when peaks with identical or near-identical mass-to-charge values appear at different retention times across batches, indicating batch-level chromatographic variation that must be corrected before alignment and annotation.

When NOT to use

  • Input is already a feature table (peak alignment matrix) with pre-corrected RTs from another tool (e.g., xcms or MZmine 2); applying redundant RT correction will introduce unnecessary noise.
  • Single-batch analysis or data from a single analytical run with no inter-batch variation; RT correction is only meaningful across distinct batches or runs.
  • Data lacks sufficient endogenous reference peaks (< 5 high-confidence, consistently detected peaks per batch), as the RT correction algorithm requires stable anchors to compute reliable shift functions.

Inputs

  • Extracted peak tables from multiple LC/HRMS batches (mzXML, mzML, or netCDF format)
  • Peak properties table with retention time (RT in minutes), m/z (mass-to-charge ratio), peak area, and intensity for each peak in each batch
  • Endogenous reference markers (list of m/z-RT pairs or automatic selection from high-intensity stable peaks)

Outputs

  • Retention-time-corrected peak table with harmonized RT values across all batches
  • Aligned peak height matrix (gap-filled and non-gap-filled) for downstream statistical analysis
  • RT correction factor per batch (drift/offset quantification for QC documentation)

How to apply

Load extracted peaks from all batches into IDSL.IPA with their retention time (RT) and m/z values. Apply the retention time correction algorithm, which uses endogenous reference markers (stable peaks present across all batches) to compute batch-specific RT shift functions. These functions are then applied to adjust all peak RTs to a common reference frame, harmonizing retention times across batches. The corrected peak table is generated with standardized RT and m/z coordinates, enabling subsequent peak alignment and annotation steps. Success hinges on the availability of sufficient endogenous reference peaks (typically 10–50 high-intensity, low-variance peaks per batch) that are consistently detected across all analytical runs.

Related tools

  • IDSL.IPA (Implements retention time correction algorithm as part of integrated peak analysis suite; executes RT harmonization across batches using endogenous reference markers) — https://github.com/idslme/IDSL.IPA
  • R (Statistical computing environment in which IDSL.IPA runs; enables custom parameter tuning and batch processing via parameter spreadsheet)
  • IDSL.UFA (Downstream molecular formula annotation tool that consumes RT-corrected and aligned peaks for compound identification) — https://github.com/idslme/IDSL.UFA
  • IDSL.UFAx (Extended molecular formula annotation tool for chemical formula assignment on RT-aligned peak tables) — https://github.com/idslme/IDSL.UFAx

Examples

library(IDSL.IPA)
IPA_workflow("path/to/IPA_parameters.xlsx")

Evaluation signals

  • RT standard deviation within each m/z tolerance window (±5 ppm) is reduced after correction compared to pre-correction data, indicating successful harmonization.
  • Peak alignment rate (percentage of peaks successfully matched across batches post-correction) is higher than pre-correction, verifiable by comparing peaklist intersection sizes.
  • Endogenous reference peak RTs converge to within ±0.5 min across all batches after correction, confirming that batch-level drift functions were estimated and applied correctly.
  • Gap-filled peak height matrix shows increased correlation between replicates (within-batch samples) compared to pre-correction, and reduced artificial clustering by batch in PCA or unsupervised clustering.
  • Output peaklists in 'peak_alignment' directory contain aligned (m/z, RT) pairs consistent with literature retention index values for known metabolites in the sample type.

Limitations

  • RT correction accuracy depends on the availability of sufficient, stable endogenous reference peaks; if reference peaks are rare or themselves subject to batch-specific variation, correction may be unreliable.
  • Non-linear RT drift (e.g., RT shift varies across the chromatographic time range) may not be fully captured by global batch-wise shift functions; localized deviations could persist.
  • Extreme batch outliers (e.g., a single run with very large RT shift or instrumental malfunction) may skew RT correction parameters if robust statistical methods are not applied; outlier batches should be identified and excluded or reanalyzed.
  • RT correction assumes comparable chromatographic conditions (temperature, column age, mobile phase composition) across batches; systematic changes in method parameters between batches may invalidate correction assumptions.

Evidence

  • [other] How does IDSL.IPA implement retention time correction to align peaks across multiple batches in LC/HRMS data?: "retention time correction across multiple batches and peak annotation"
  • [other] Load extracted peaks from multiple batches with their retention time and mass-to-charge values into IDSL.IPA.: "Load extracted peaks from multiple batches with their retention time and mass-to-charge values into IDSL.IPA"
  • [other] Apply retention time correction algorithm across batches to align peaks with similar m/z and RT profiles.: "Apply retention time correction algorithm across batches to align peaks with similar m/z and RT profiles"
  • [readme] IDSL.IPA is a suite of new algorithms covering extracted ion chromatogram (EIC) candidate generation, peak detection, peak property evaluation, recursive mass correction, retention time correction across multiple batches and peak annotation.: "suite of new algorithms covering extracted ion chromatogram (EIC) candidate generation, peak detection, peak property evaluation, recursive mass correction, retention time correction across multiple"
  • [readme] Retention time correction using endogenous reference markers for multi-batch large scale studies: "Retention time correction using endogenous reference markers for multi-batch large scale studies"
  • [readme] extracts peaks for organic small molecules from untargeted liquid chromatography high resolution mass spectrometry (LC/HRMS) data in population scale projects: "extracts peaks for organic small molecules from untargeted liquid chromatography high resolution mass spectrometry (LC/HRMS) data in population scale projects"