Ccs calibration polynomial fitting

Skill HolobiomicsLab/asb-skill-collections/collections/metabolomics/v2/skills/ccs-calibration-polynomial-fitting

Use when you have acquired tunemix or reference standard data in ion mobility spectrometry with known m/z, drift time, and CCS values, and you need to establish a drift-time-to-CCS mapping for a specific instrument, ionization mode (positive or negative), and buffer gas.From its SKILL.md

Install
npx -y skills add HolobiomicsLab/asb-skill-collections --skill ccs-calibration-polynomial-fitting

Assembled from the repository path, not quoted from the project. Check it against their README if it does not work.

One thing to look at

  • 15 stars15 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.

What its file declares

Copied from the file, not written here

The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.

SKILL.md

7.8 KB, ~1.3k tokens by cl100k_base, as published. Nobody here has run it

ccs-calibration-polynomial-fitting

Summary

Fit a polynomial calibration model to convert ion mobility drift times and reference m/z values into collision cross section (CCS) coefficients using the single-field calibration equation. This skill produces calibration coefficients (beta and tfix) that enable downstream CCS prediction for unknown ions in positive- or negative-mode ion mobility spectrometry.

When to use

Apply this skill when you have acquired tunemix or reference standard data in ion mobility spectrometry with known m/z, drift time, and CCS values, and you need to establish a drift-time-to-CCS mapping for a specific instrument, ionization mode (positive or negative), and buffer gas. The skill is necessary as the first step in any CCS-based feature characterization workflow.

When NOT to use

  • Do not use this skill if reference CCS standards are not available or not accurately characterized; the calibration will be unreliable.
  • Do not use this skill if drift time and m/z data have not been properly preprocessed (e.g., baseline-subtracted, aligned) or contain significant instrumental noise; garbage input produces garbage calibration.
  • Do not use this skill if switching between ionization modes or buffer gases without recalibrating; calibration coefficients are mode- and gas-specific.

Inputs

  • tunemix reference m/z values (numpy array or pandas Series)
  • tunemix experimental drift times (numpy array or pandas Series, units: milliseconds)
  • known CCS values for reference standards (numpy array, units: Ų)
  • charge states (integer array or scalar)
  • buffer gas mass (float, atomic mass units)
  • calibration mode identifier (string: 'positive' or 'negative')

Outputs

  • calibration coefficient beta (float)
  • calibration coefficient tfix (float)
  • R-squared goodness-of-fit metric (float, range 0–1)
  • fitted calibration model object (DEIMoS calibration model)

How to apply

Load reference tunemix data (m/z, drift_time, known CCS values, charge states) from an HDF5 file using deimos.load(). Call deimos.calibration.calibrate_ccs() with the reference m/z and drift_time arrays, specifying the calibration mode (positive or negative), known CCS standards, and buffer gas mass. The function applies the single-field calibration equation (Stow et al. 2017) to solve for calibration coefficients beta and tfix. Extract the returned R-squared metric and polynomial coefficients. Validate that the R-squared value meets the threshold of ≥0.9999; values below this indicate poor model fit and may signal data quality issues, incorrect reference standards, or instrument miscalibration.

Related tools

  • DEIMoS (Python API providing calibrate_ccs() function and load() utility for HDF5 file I/O; applies single-field calibration equation to compute beta and tfix coefficients) — https://github.com/pnnl/deimos
  • numpy (Numerical array operations for m/z, drift_time, and CCS value handling during calibration)
  • Python (Execution environment and scripting language for DEIMoS API calls and calibration workflow)

Examples

tune_pos = deimos.load('example_tune_pos.h5', key='ms1'); calibration = deimos.calibration.calibrate_ccs(tune_pos['mz'].values, tune_pos['drift_time'].values, known_ccs, charges, buffer_gas_mass, mode='positive')

Evaluation signals

  • R-squared metric ≥ 0.9999 (threshold from paper validation examples; values <0.99 indicate poor fit)
  • Returned beta and tfix coefficients are finite, positive numbers and match expected physical ranges for the instrument and buffer gas
  • Calibration model can be applied to independent test m/z and drift_time pairs and produce CCS predictions within ±5% of known standard values (typical IMS accuracy)
  • Residual plot of predicted vs. observed drift times shows random scatter around zero with no systematic bias across m/z or drift_time ranges
  • Model coefficients are reproducible when recalibrated with the same reference data (deterministic function)

Limitations

  • Calibration is specific to a single ionization mode (positive or negative) and buffer gas; switching modes requires recalibration.
  • Accuracy depends critically on the quality and accuracy of reference CCS standards; contaminated or mislabeled standards will corrupt the model.
  • The single-field calibration equation assumes ideal gas behavior and may not account for instrumental nonlinearities or space-charge effects at high ion currents.
  • Extrapolation beyond the m/z and drift_time ranges covered by reference standards is not reliable; predictions are most trustworthy within the calibration domain.

Evidence

  • [other] The calibrate_ccs routine accepts reference m/z values, known CCS values, charge states, experimental drift times, and buffer gas mass, then applies the single-field calibration equation to compute calibration coefficients beta and tfix: "The calibrate_ccs routine accepts reference m/z values, known CCS values, charge states, experimental drift times, and buffer gas mass, then applies the single-field calibration equation to compute"
  • [other] yielding calibration model fit with r-squared of 0.9999784552958134: "yielding calibration model fit with r-squared of 0.9999784552958134"
  • [other] Load tunemix positive-mode reference data (m/z, drift_time, known CCS values) from example_tune_pos.h5 using deimos.load(): "Load tunemix positive-mode reference data (m/z, drift_time, known CCS values) from example_tune_pos.h5 using deimos.load()"
  • [other] confirm the reported R-squared is ≥0.9999 (threshold criterion from paper examples): "confirm the reported R-squared is ≥0.9999 (threshold criterion from paper examples)"
  • [readme] Functionality includes feature detection, feature alignment, collision cross section (CCS) calibration, isotope detection, and MS/MS spectral deconvolution: "Functionality includes feature detection, feature alignment, collision cross section (CCS) calibration, isotope detection, and MS/MS spectral deconvolution"
  • [readme] algorithm implementations simultaneously utilize all dimensions to (i) offer greater separation between features, thus improving detection sensitivity: "algorithm implementations simultaneously utilize all dimensions to (i) offer greater separation between features, thus improving detection sensitivity"

What ships with it

Read from the repository

Just SKILL.md. No reference files, no scripts.

Keep looking

Skills are one crate of 325,949. Ordering is by how many stacks a row turns up in, so the top of any crate is what has actually been picked rather than what has the most stars.