External calibration model fitting

Skill HolobiomicsLab/asb-skill-collections/collections/metabolomics/v2/skills/external-calibration-model-fitting

Use when you have acquired targeted mass spectrometry data with measured ion intensities for known standard compounds at multiple concentration levels, and you need to convert sample intensities into absolute or relative concentrations.From its SKILL.md

Install
npx -y skills add HolobiomicsLab/asb-skill-collections --skill external-calibration-model-fitting

Assembled from the repository path, not quoted from the project. Check it against their README if it does not work.

One thing to look at

  • 15 stars15 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.

What its file declares

Copied from the file, not written here

The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.

SKILL.md

7.3 KB, ~1.1k tokens by cl100k_base, as published. Nobody here has run it

external-calibration-model-fitting

Summary

Fit a calibration model (linear or non-linear regression) to relate mass spectrometry signal intensities to known standard concentrations, enabling conversion of sample intensities into quantitative predictions. This skill addresses intensity drift in MS datasets through multiple correction strategies (internal standard, drift correction, bracketing, weighted bracketing).

When to use

You have acquired targeted mass spectrometry data with measured ion intensities for known standard compounds at multiple concentration levels, and you need to convert sample intensities into absolute or relative concentrations. Apply this skill especially when intensity drift is observed across the measurement sequence—drift that would bias uncorrected intensity-to-concentration relationships.

When NOT to use

  • Input is already a processed feature table or normalized concentration matrix—this skill operates on raw or drift-corrected intensities, not pre-quantified data.
  • No standard compounds with known concentrations are available—the skill requires a calibration set to fit the model.
  • Measurement sequence order is unknown or unavailable—drift correction strategies (bracketing, IS correction) depend on knowing sample position and timing.

Inputs

  • Standard compound MS intensity measurements (numeric intensities per compound)
  • Standard compound reference concentrations (known concentration values)
  • Sample MS intensity data (measured intensities for unknown samples)
  • Measurement sequence metadata (sample order, time stamps for drift detection)

Outputs

  • Fitted calibration model (linear or quadratic regression parameters)
  • Model diagnostics (R², residuals, weights applied per standard level)
  • Sample concentration predictions (predicted concentrations in tabular format)
  • Drift-corrected intensities (after internal standard, bracketing, or statistical correction)

How to apply

Load standard compound MS intensity measurements alongside their known concentrations into QuantyFey. Select a regression model type (linear or quadratic) and choose appropriate standard levels; the automatic optimization module can assist model selection. Apply one or more drift correction strategies: internal standard (IS) correction for sample-by-sample normalization, statistical drift correction for systematic instrument drift, custom bracketing to assign calibration data to predefined measurement blocks, or weighted bracketing to weight calibration curves by sample position between calibrations. Fit the model to the corrected intensities. Validate the fitted model by examining residuals and R² values across the calibration range. Once validated, apply the fitted model to sample intensities to generate predicted concentrations.

Related tools

  • QuantyFey (Shiny application that implements external calibration model fitting with drift correction, regression optimization (linear/quadratic), and bracketing-based quantification strategies) — https://github.com/CDLMarkus/QuantyFey

Evaluation signals

  • Fitted model R² is ≥ 0.95 across the calibration range, indicating good linearity or quadratic fit.
  • Residuals are randomly distributed around zero with no systematic trends across concentration levels or measurement sequence.
  • Applied drift correction strategy (IS, statistical, bracketing) reduces intensity variance between replicate standards measured at different times.
  • Sample concentrations fall within the interpolation range of the calibration curve (not extrapolated far beyond standard range).
  • Exported concentration predictions are in consistent tabular format with identifiable sample IDs and concentration units matching the standard reference data.

Limitations

  • QuantyFey standalone version is compatible with Windows operating systems only; Linux users must use the Apptainer version, and macOS setup is slow and difficult.
  • The tool does not provide raw mass spectrometry data integration; it accepts pre-extracted intensity data and operates as a post-acquisition quantification module.
  • Automatic model optimization selects between linear and quadratic models only; more complex non-linear models require manual parameter adjustment.
  • Bracketing and weighted bracketing strategies assume calibration standards are interspersed regularly throughout the measurement sequence; performance degrades if standards are clustered.
  • No changelog is available, limiting transparency on version differences and potential bug fixes.

Evidence

  • [other] Fit a calibration model (linear or non-linear regression) relating intensity to concentration using the standards.: "Fit a calibration model (linear or non-linear regression) relating intensity to concentration using the standards."
  • [other] Load standard compound MS intensity measurements and their known concentrations.: "Load standard compound MS intensity measurements and their known concentrations."
  • [other] Apply the fitted calibration model to convert sample intensities into predicted concentrations.: "Apply the fitted calibration model to convert sample intensities into predicted concentrations."
  • [readme] It is specifically designed to address intensity drifts in datasets, offering multiple correction strategies: "It is specifically designed to address intensity drifts in datasets, offering multiple correction strategies"
  • [readme] Interactive Regression Model Optimization - manual adjustment of model, weights, standard levels etc.: "Interactive Regression Model Optimization - manual adjustment of model, weights, standard levels etc."
  • [readme] automatic selection of linear or quadratic regression model, and selection of appropriate standards: "automatic selection of linear or quadratic regression model, and selection of appropriate standards"
  • [readme] Internal Standard (IS) correction, Drift Correction using statistical models, Custom Bracketing Quantification, Weighted Bracketing: "Internal Standard (IS) correction, Drift Correction using statistical models, Custom Bracketing Quantification, Weighted Bracketing"

What ships with it

Read from the repository

Just SKILL.md. No reference files, no scripts.