Signal apodization configuration

Skill HolobiomicsLab/asb-skill-collections/collections/metabolomics/v2/skills/signal-apodization-configuration

Curated, evidence-grounded skill and software-tool collections for scientific AI agents, generated by the AgenticScienceBuilder

Install
npx -y skills add HolobiomicsLab/asb-skill-collections --skill signal-apodization-configuration

Assembled from the repository path, not quoted from the project. Check it against their README if it does not work.

One thing to look at

  • 14 stars14 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.

What its author says it does

Copied from the file, not written here

Use when when processing raw Bruker Solarix transient files (.d format) destined for FT-MS analysis, especially for ESI-negative or low-abundance natural organic matter samples where baseline noise and side-lobe artifacts around intense peaks degrade peak picking and formula assignment accuracy.

The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.

SKILL.md

7.3 KB, ~1.3k tokens by cl100k_base, as published. Nobody here has run it

Signal Apodization Configuration

Summary

Configure and apply apodization window functions (e.g., Hanning) to FT-ICR transient data prior to magnitude-mode Fourier transformation to reduce spectral artifacts and improve mass peak definition. Apodization trades resolving power for dynamic range and baseline quality.

When to use

When processing raw Bruker Solarix transient files (.d format) destined for FT-MS analysis, especially for ESI-negative or low-abundance natural organic matter samples where baseline noise and side-lobe artifacts around intense peaks degrade peak picking and formula assignment accuracy.

When NOT to use

  • Input is already a centroided or processed mass list (.csv, .xlsx, or mzML); apodization applies only to raw transients.
  • Analysis requires maximum resolving power (e.g., sub-ppm mass accuracy for ultra-high-mass lipids); Hanning apodization sacrifices ~30% resolving power.
  • Transient data is already apodized by the instrument acquisition software; double-apodization will over-damp the signal.

Inputs

  • Bruker Solarix transient file (.d directory)
  • Transient processing parameters object (MSParameters)
  • Raw time-domain transient data (ser or fid files within .d)

Outputs

  • Apodized transient (time-domain signal after windowing)
  • Mass spectrum object with detected m/z peaks and abundances
  • Peak list (m/z, intensity, resolving power per peak)

How to apply

After loading a Bruker Solarix transient (.d file) using CoreMS's ReadBrukerSolarix class, configure the transient processing parameters to select and apply a Hanning apodization window function before zero-filling and FFT. Hanning apodization suppresses spectral ringing by smoothing the transient edges, reducing side-lobe intensity at the cost of ~30% resolving power reduction. Set the apodization window parameter in the MSParameters before calling the process method to generate the magnitude-mode mass spectrum. The choice of Hanning (vs. other windows like Blackman or Kaiser) depends on the trade-off desired: Hanning offers moderate side-lobe suppression and is widely used for ESI-FT workflows. Verify output by inspecting the resulting m/z peak count, mass accuracy distribution, and visual baseline quality in matplotlib plots.

Related tools

  • CoreMS (Provides ReadBrukerSolarix class to load and configure transient processing with apodization window selection; MSParameters object stores and applies apodization settings during FFT pipeline.) — https://github.com/EMSL-Computing/CoreMS
  • Bruker Solarix instrument (Source of raw transient data (.d file format); metadata includes calibration constants and acquisition parameters needed for frequency-to-m/z conversion post-FFT.)
  • numpy (Underlying numerical library for windowing function calculation and FFT magnitude computation.)
  • matplotlib (Visualization of apodized vs. non-apodized spectra for quality assessment and baseline inspection.)

Examples

from corems.transient.input.brukerSolarix import ReadBrukerSolarix; lcms = ReadBrukerSolarix('ESI_NEG_SRFA.d'); lcms.apodization_method = 'Hanning'; lcms.zero_fill_factor = 1; lcms.process()

Evaluation signals

  • Output mass spectrum peak count and m/z range match expected analyte composition (e.g., ESI_NEG_SRFA: 10588 detected m/z peaks from m/z 155.87–999.66).
  • Baseline noise floor is flat and free of significant spectral ringing artifacts around intense peaks.
  • Mass accuracy (calculated m/z vs. theoretical) remains within calibration tolerance (typically <2 ppm for FT-ICR).
  • Peak width (full width at half maximum) and resolving power are consistent with the known magnetic field strength (e.g., 15 Tesla) and transient collection time.
  • Comparison of peak abundances before/after apodization shows expected redistribution (lower peak height, wider baseline, reduced artifacts).

Limitations

  • Hanning apodization reduces resolving power by ~30%, making it unsuitable for very high-mass or isobaric analytes requiring sub-ppm separation.
  • Choice of window function (Hanning vs. Blackman vs. Kaiser) is problem-dependent and not automatically optimized; manual parameter sweep may be needed for novel sample types.
  • Transient truncation or incomplete data collection will degrade apodization effectiveness; quality depends on signal-to-noise and transient length.
  • Double-apodization (instrument + software) can over-suppress signal; user must verify whether the acquisition already applied windowing.

Evidence

  • [other] Using Hanning apodization with zero fill settings on ESI_NEG_SRFA.d produces a mass spectrum with 10588 detected m/z peaks spanning from m/z 155.87 to m/z 999.66.: "Using Hanning apodization with zero fill settings on ESI_NEG_SRFA.d produces a mass spectrum with 10588 detected m/z peaks spanning from m/z 155.87 to m/z 999.66."
  • [other] Import the ReadBrukerSolarix class from CoreMS and load the ESI_NEG_SRFA.d transient file. Configure transient processing parameters to apply Hanning apodization window function. Process the transient data to generate the mass spectrum with averaging applied.: "Import the ReadBrukerSolarix class from CoreMS and load the ESI_NEG_SRFA.d transient file. Configure transient processing parameters to apply Hanning apodization window function. Process the"
  • [readme] Apodization, Zerofilling, and Magnitude mode FT: "Apodization, Zerofilling, and Magnitude mode FT"
  • [readme] from corems.transient.input.brukerSolarix import ReadBrukerSolarix: "from corems.transient.input.brukerSolarix import ReadBrukerSolarix"
  • [readme] The data structures were designed with an intuitive, mass spectrometric hierarchical structure, thus allowing organized and easy access to the data and calculations.: "The data structures were designed with an intuitive, mass spectrometric hierarchical structure, thus allowing organized and easy access to the data and calculations."