Y ion signal detection and noise assessment

Skill HolobiomicsLab/asb-skill-collections/collections/metabolomics/v1/skills/y-ion-signal-detection-and-noise-assessment

Use when you have extracted a centroided MS/MS spectrum from a Thermo Orbitrap raw file (via rawrr::readSpectrum or equivalent) and need to verify that the observed y-ion fragments for a known peptide precursor exhibit signal-to-noise ratios consistent with high-quality fragmentation.From its SKILL.md

Install
npx -y skills add HolobiomicsLab/asb-skill-collections --skill y-ion-signal-detection-and-noise-assessment

Assembled from the repository path, not quoted from the project. Check it against their README if it does not work.

One thing to look at

  • 15 stars15 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.

What its file declares

Copied from the file, not written here

The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.

SKILL.md

8.3 KB, ~1.6k tokens by cl100k_base, as published. Nobody here has run it

y-ion signal detection and noise assessment

Summary

Detects and validates y-ion fragment signals in centroided tandem MS spectra by comparing peak intensities to local baseline noise, ensuring peptide fragmentation quality meets analytical thresholds. Applied to high-resolution Orbitrap data to verify that observed ion signals are tens to hundreds of times above the noise floor.

When to use

You have extracted a centroided MS/MS spectrum from a Thermo Orbitrap raw file (via rawrr::readSpectrum or equivalent) and need to verify that the observed y-ion fragments for a known peptide precursor exhibit signal-to-noise ratios consistent with high-quality fragmentation. Use this skill when evaluating spectral quality, validating peptide identification confidence, or assessing whether an LC-MS run meets performance benchmarks for a targeted peptide.

When NOT to use

  • Input spectrum is profile-mode (not centroided); baseline noise estimation and peak matching will be unreliable without centroiding preprocessing.
  • Peptide sequence is unknown or ambiguous; theoretical y-fragment m/z calculation depends on accurate sequence annotation.
  • Raw file is from a non-Orbitrap or non-high-resolution instrument where baseline noise characteristics differ substantially from Orbitrap noise profiles.

Inputs

  • centroided MS/MS spectrum (m/z and intensity arrays) from rawrr::readSpectrum or equivalent
  • peptide sequence (e.g., LGGNEQVTR)
  • precursor m/z and charge state
  • instrument metadata (resolving power, AGC target, injection time)
  • mass tolerance threshold (e.g., 10 ppm)

Outputs

  • identified y-ion peak positions and intensities
  • signal-to-noise ratios for each y-ion
  • local baseline noise estimates
  • spectral quality assessment (pass/fail based on S/N thresholds)
  • AGC injection efficiency metric (actual / maximum injection time %)

How to apply

Extract the m/z and intensity arrays from the centroided spectrum object returned by rawrr::readSpectrum. Calculate theoretical y-fragment m/z values for the peptide sequence (e.g., LGGNEQVTR as doubly charged 487.2567 m/z). Identify observed y-ion signals by matching experimental m/z to theoretical m/z with appropriate tolerance (e.g., 10 ppm for Orbitrap resolving power 30,000 at 200 m/z). For each y-ion peak, measure the signal intensity and estimate local baseline noise from adjacent spectral regions. Calculate the signal-to-noise ratio as peak intensity divided by local noise floor. Verify that all y-ions exceed tens to hundreds counts above the noise level to confirm high spectral quality. Document the AGC injection time (e.g., 2.8 ms of 55 ms maximum) as a metric of ion accumulation efficiency.

Related tools

Examples

S <- rawrr::readSpectrum(rawfile = '20181113_010_autoQC01.raw', scan = 9594); y_ions <- c(175.119, 288.203, 403.230); observed_mz <- S[[1]]$mz; observed_int <- S[[1]]$intensity; sapply(y_ions, function(y) min(abs(observed_mz - y)) < 0.01)

Evaluation signals

  • All identified y-ions have signal intensity ≥ tens to hundreds counts above measured local baseline noise (quantitative S/N assessment).
  • Observed y-ion m/z values match theoretical m/z within specified tolerance (e.g., 10 ppm) and fragment ladder is continuous or mostly continuous.
  • AGC injection time is ≤ 55 ms maximum for Orbitrap and represents typical efficiency (e.g., ~5% in the example indicates rapid ion accumulation).
  • Absence of unexplained gaps in the y-ion series; missing y-ions are rare or attributable to known loss patterns (e.g., loss of ammonia or water).
  • Baseline noise floor is stable across the m/z range and does not show artifacts (e.g., electronic noise spikes) that would inflate apparent S/N.

Limitations

  • Noise estimation depends on the definition of 'local baseline'; adjacent spectral regions must be carefully selected to avoid contamination from nearby peaks or spectral artifacts.
  • Signal-to-noise assessment assumes centroided data; profile-mode spectra require peak-picking preprocessing and may yield different noise estimates.
  • Theoretical y-fragment m/z calculation assumes the peptide sequence is correct and charge state is known; sequence ambiguity or charge miscalculation will cause false negatives.
  • Windows systems require decimal symbol configured as '.' for proper rawrr data extraction; misconfiguration will cause data parsing failures.
  • The skill does not account for variable ionization efficiency across different y-ion m/z ranges; hydrophobic or highly charged fragments may show anomalously low or high intensities independent of instrumental performance.

Evidence

  • [other] Identify y-ion signals by matching observed m/z values to theoretical y-fragment m/z of the precursor peptide (LGGNEQVTR/2). Calculate signal-to-noise ratio for each y-ion by comparing peak intensity to local baseline noise and verify all y-ions exceed tens to hundreds counts above noise floor.: "Identify y-ion signals by matching observed m/z values to theoretical y-fragment m/z of the precursor peptide (LGGNEQVTR/2). Calculate signal-to-noise ratio for each y-ion by comparing peak intensity"
  • [other] all y-ion signals for LGGNEQVTR++ peptide are several tens to hundreds of times above the noise level, demonstrating high spectral quality.: "all y-ion signals for LGGNEQVTR++ peptide are several tens to hundreds of times above the noise level, demonstrating high spectral quality."
  • [methods] Specifically, R functions requesting access to data stored in binary raw files (reader family functions listed in Table 1) invoke compiled C# wrapper methods using a system call: "invoke compiled C# wrapper methods using a system call"
  • [results] the C-trap managed to collect the defined 100,000 charges within 2.8 ms, corresponding to only ~r format((2.8/55)*100, digits = 1)% of the maximum injection time of 55 ms: "the C-trap managed to collect the defined 100,000 charges within 2.8 ms, corresponding to only ~5% of the maximum injection time of 55 ms"
  • [methods] The example file 20181113_010_autoQC01.raw used throughout this manuscript contains Fourier-transformed Orbitrap spectra (FTMS) recorded on a Thermo Fisher Scientific Q Exactive HF: "Fourier-transformed Orbitrap spectra (FTMS) recorded on a Thermo Fisher Scientific Q Exactive HF"