Mass error calculation and validation
Curated, evidence-grounded skill and software-tool collections for scientific AI agents, generated by the AgenticScienceBuilder
npx -y skills add HolobiomicsLab/asb-skill-collections --skill mass-error-calculation-and-validationAssembled from the repository path, not quoted from the project. Check it against their README if it does not work.
One thing to look at
- 14 stars14 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.
What its author says it does
Copied from the file, not written here
Use when when annotating observed mass spectrometry peaks against theoretical fragment ions (b, y, or other ion types) using ProForma 2.0 peptidoforms, compute the m/z deviation for each matched peak to verify that the annotation adheres to your specified mass tolerance (e.g., ±10 ppm or ±0.
The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.
SKILL.md
7.5 KB, ~1.4k tokens by cl100k_base, as published. Nobody here has run it
mass-error-calculation-and-validation
Summary
Calculate and validate the mass deviation (Δm/z) between theoretical fragment ion m/z values and observed spectrum peaks to assess annotation accuracy and fragment ion assignment correctness. This skill ensures that annotated ions fall within specified mass tolerance thresholds, a prerequisite for reliable peptidoform identification.
When to use
When annotating observed mass spectrometry peaks against theoretical fragment ions (b, y, or other ion types) using ProForma 2.0 peptidoforms, compute the m/z deviation for each matched peak to verify that the annotation adheres to your specified mass tolerance (e.g., ±10 ppm or ±0.02 Da) and to quantify annotation reliability. Use this skill after calling annotate_proforma() to validate that each peak's assigned ion type, charge state, and m/z are consistent with experimental precision.
When NOT to use
- The observed spectrum has not been quality-filtered (e.g., precursor peak and low-intensity noise not removed); preprocessing should precede annotation validation.
- Fragment mass tolerance parameters have not been empirically determined for your instrument and data format; validation will be arbitrary if tolerance is misspecified.
- The peptide string is not in ProForma 2.0 format or contains unrecognized modifications not defined in Unimod or PSI-MOD; annotation will fail or produce nonsensical m/z values.
Inputs
- ProForma 2.0 peptide string (e.g., 'DLTDYLM[Oxidation]K')
- Observed mass spectrum (MsmsSpectrum object with peak m/z and intensity)
- Theoretical fragment ion m/z values (computed from peptide and modification positions)
- Fragment mass tolerance parameters (fragment_tol_mass and fragment_tol_mode)
Outputs
- Annotated spectrum peaks with ion type, charge state, and m/z deviation
- List of validated (matched) peaks and their mass errors
- Flag indicators for peaks exceeding mass tolerance
- Summary statistics of mass error distribution (mean, std dev, range)
How to apply
After generating theoretical b and y ion m/z values for a ProForma 2.0 peptide string using spectrum_utils.fragment_annotation, match each observed spectrum peak to its nearest theoretical m/z within the specified fragment_tol_mass and fragment_tol_mode (e.g., 'Da' or 'ppm'). For each matched peak, compute the absolute and relative mass error: absolute error = |observed_mz - theoretical_mz| and relative error (ppm) = (observed_mz - theoretical_mz) / theoretical_mz × 1e6. Validate that all errors remain within the tolerance threshold; peaks exceeding the tolerance should be flagged as potential misassignments. Iterate through the annotated peaks and inspect the m/z deviation to identify systematic biases (e.g., calibration drift) or outliers that may indicate incorrect ion type or charge assignment.
Related tools
- spectrum_utils (Provides the MsmsSpectrum class and annotate_proforma() method to compute theoretical fragment ions and match them to observed peaks; enables extraction of m/z deviations for validation.) — https://github.com/bittremieux/spectrum_utils/
- ProForma 2.0 (Standard notation for encoding peptide sequences and post-translational modifications; spectrum_utils uses ProForma to parse peptide strings and retrieve modification masses from Unimod.) — https://www.psidev.info/proforma
- Unimod (Controlled vocabulary of protein modifications with accurate masses; spectrum_utils resolves modification names (e.g., 'Oxidation') to masses for theoretical m/z calculation.) — https://www.unimod.org/
Examples
spectrum.annotate_proforma('DLTDYLM[Oxidation]K', fragment_tol_mass=0.02, fragment_tol_mode='Da', ion_types='by'); [print(f"{peak.ion_type}{peak.charge}: {peak.mz:.4f} (Δ={peak.mz - theoretical_mz:.6f} Da)") for peak in spectrum.annotated_peaks]
Evaluation signals
- All annotated peaks have m/z deviations within the specified fragment_tol_mass and fragment_tol_mode (e.g., |Δm/z| < 0.02 Da or < 10 ppm).
- Mass error distribution is centered near zero with no systematic drift; median and mean deviations should be close and symmetric around zero.
- Ion type, charge state, and m/z deviation are internally consistent (e.g., y^2+ ions have lower m/z than y^1+, and higher-charge ions show proportionally tighter errors).
- Peaks flagged as exceeding tolerance represent <5% of total annotated peaks (unless instrument calibration is poor, in which case recalibration is warranted).
- Comparing mass errors across multiple spectra from the same sample shows reproducible patterns, indicating systematic instrument behavior rather than random noise.
Limitations
- Mass error calculation assumes accurate fragment_tol_mass and fragment_tol_mode parameters; if these are misspecified, validation will incorrectly accept or reject peaks.
- The method does not account for isotope patterns or in-source fragmentation; peaks from 13C isotopologues or loss species will exhibit larger m/z deviations and may be misassigned.
- Peptides with complex or non-canonical modifications not present in Unimod or PSI-MOD cannot be accurately annotated, leading to spurious m/z deviations.
- No changelog or version history is available for spectrum_utils, making it difficult to track changes in mass error calculation logic across releases.
Evidence
- [other] compute theoretical b and y ion m/z values and match them to observed spectrum peaks within the specified tolerance: "Call MsmsSpectrum.annotate_proforma(peptide, fragment_tol_mass, fragment_tol_mode, ion_types='by') to compute theoretical b and y ion m/z values and match them to observed spectrum peaks within the"
- [other] each peak's assigned ion_type, charge state, and m/z deviation from theoretical mass are correct and consistent: "Iterate through annotated peaks and validate that each peak's assigned ion_type ('b' or 'y'), charge state, and m/z deviation from theoretical mass are correct and consistent with the input"
- [other] annotate_proforma enables ion type annotation based on ProForma 2.0 specification: "Annotating observed spectrum fragments using the ProForma 2.0 specification for (modified) peptidoforms."
- [other] fragment ions can be annotated based on ProForma 2.0 specification: "fragment ions can be annotated based on the ProForma 2.0 specification"
- [other] Modifications are defined by controlled vocabularies including Unimod: "Modifications are defined by controlled vocabularies (CVs), including Unimod"