Mass spectrum calibration validation

Skill HolobiomicsLab/asb-skill-collections/collections/metabolomics/v1/skills/mass-spectrum-calibration-validation

Curated, evidence-grounded skill and software-tool collections for scientific AI agents, generated by the AgenticScienceBuilder

Install
npx -y skills add HolobiomicsLab/asb-skill-collections --skill mass-spectrum-calibration-validation

Assembled from the repository path, not quoted from the project. Check it against their README if it does not work.

One thing to look at

  • 14 stars14 stars. Stars are a popularity signal and not a quality one, but at this level it is likely that nobody has read this closely except its author, and you would be relying on your own review.

What its author says it does

Copied from the file, not written here

Use when after applying frequency domain calibration (Ledford, linear, or quadratic equation) to a raw FT-ICR mass spectrum, validate the calibration quality by measuring residual mass errors across the m/z range.

The file declares its own license as CC-BY-4.0. That is the author’s claim about this one file, and it is not the same thing as the license GitHub reports for the repository, which is listed with the other numbers below.

SKILL.md

7.2 KB, ~1.3k tokens by cl100k_base, as published. Nobody here has run it

mass-spectrum-calibration-validation

Summary

Validate calibrated mass spectra by comparing theoretical and measured m/z values across multiple calibration equations and assessing mass error distributions. This skill ensures that frequency-to-m/z domain transformations have been applied correctly before molecular formula assignment.

When to use

After applying frequency domain calibration (Ledford, linear, or quadratic equation) to a raw FT-ICR mass spectrum, validate the calibration quality by measuring residual mass errors across the m/z range. Use this skill when preparing spectra for downstream molecular formula search, especially when first-hit or all-hits assignment modes will be compared, since poor calibration can artificially inflate or suppress assignment counts.

When NOT to use

  • Input spectrum has not yet undergone any calibration — use calibration assignment first.
  • Spectrum is centroid-only without access to peak position metadata — validation requires measured vs. theoretical m/z pairs.
  • Analysis goal is exploratory and tolerates large mass errors (e.g., compound class screening rather than molecular formula assignment).

Inputs

  • Processed mass spectrum (Bruker CompassXtract, Bruker transients, ThermoFisher .raw, CoreMS .hdf5)
  • Calibration model parameters (Ledford, linear, or quadratic coefficients)
  • Reference mass list or internal standard compounds with known accurate m/z

Outputs

  • Mass error distribution table (m/z, measured m/z, calculated m/z, error in ppm or Da, residuals)
  • Summary statistics (mean, median, std, min, max of mass error)
  • Calibration validation plots (mass error vs. m/z, error histogram, Q-Q plot)
  • Pass/fail assessment and recommendation for recalibration if needed

How to apply

Load the processed mass spectrum from vendor-specific formats (Bruker .d, ThermoFisher .raw, or CoreMS .hdf5) and extract both measured and theoretical m/z values. Calculate mass error (ppm or absolute m/z difference) for each detected peak using the calibration equation applied. Compute summary statistics (mean, median, standard deviation, min, max) of the mass error distribution. Generate plots (e.g., mass error vs. m/z, histogram of error distribution) to identify systematic bias or outliers. Compare error metrics against acceptance thresholds appropriate to your analysis (typically < 2–5 ppm for FT-ICR). If residual errors exceed tolerance, re-calibrate using alternative reference compounds or a different calibration model (e.g., switch from linear to quadratic) before proceeding to formula assignment.

Related tools

  • CoreMS (Framework for loading spectrum data, applying calibration models (Ledford, linear, quadratic equations), and computing mass error metrics) — https://github.com/EMSL-Computing/CoreMS
  • pandas (Tabulating and summarizing mass error statistics (mean, median, std, min, max) into comparison tables)
  • numpy (Computing mass error arrays and statistical distributions)
  • matplotlib (Generating diagnostic plots (mass error vs. m/z, error histograms, calibration curves))

Examples

from corems.encapsulation.factory.parameters import MSParameters; spectrum.calibrate(freq_shift=0); errors = [(p.mz_theor - p.mz) / p.mz * 1e6 for p in spectrum]; print(f'Mean error: {np.mean(errors):.3f} ppm, Std: {np.std(errors):.3f} ppm')

Evaluation signals

  • Mass error mean is centered near zero ppm and does not show systematic bias across the m/z range (slope close to 0).
  • Mass error standard deviation is within specification for the instrument (typically < 2–5 ppm for FT-ICR).
  • Residual mass error histogram is approximately normal-distributed, indicating random rather than systematic error.
  • No outliers or anomalous peaks with error > 3σ from the mean (indicates miscalibrated peaks or contaminants).
  • Recalibration with an alternative model (e.g., quadratic vs. linear) does not substantially reduce mean absolute error, confirming convergence.

Limitations

  • Calibration validation requires high-confidence reference masses; sparse or low-quality reference lists can produce unreliable error estimates.
  • Mass error assessment is local to the m/z range covered by detected peaks; extrapolation beyond this range is not validated.
  • Different calibration models (Ledford, linear, quadratic) may show different residual error profiles; the choice of model affects interpretation of validation results.
  • The CoreMS README notes that calibration functions support frequency and m/z domain transformations but does not specify automatic model selection; users must choose the appropriate equation for their instrument and sample type.

Evidence

  • [readme] Frequency and m/z domain calibration functions: LedFord equation, Linear equation, Quadratic equation: "Frequency and m/z domain calibration functions: - LedFord equation - Linear equation - Quadratic equation"
  • [other] Extract and tabulate assignment counts and score distribution statistics (mean, median, std, min, max) for each mode.: "Extract and tabulate assignment counts and score distribution statistics (mean, median, std, min, max) for each mode."
  • [readme] CoreMS supports direct access for almost all vendors' data formats, allowing for the centralization and automation of all data processing workflows: "CoreMS supports direct access for almost all vendors' data formats, allowing for the centralization and automation of all data processing workflows from the raw signal to data annotation and curation."
  • [other] Load the calibrated mass spectrum from ESI_NEG_SRFA.d dataset and reference file SRFA.ref.: "Load the calibrated mass spectrum from ESI_NEG_SRFA.d dataset and reference file SRFA.ref."
  • [readme] High Resolution Mass Spectrum Simulations: Peak shape (Lorentz, Gaussian, Voigt, and pseudo-Voigt), Peak fitting for peak shape definition, Peak position in function of data points, signal to noise and resolving power: "High Resolution Mass Spectrum Simulations: Peak shape (Lorentz, Gaussian, Voigt, and pseudo-Voigt), Peak fitting for peak shape definition, Peak position in function of data points, signal to noise"

What ships with it

Read from the repository

Just SKILL.md. No reference files, no scripts.